Rt‐15 flow monitor – AW Gear Meters RT-15 User Manual
Page 17
![background image](/manuals/366492/17/background.png)
RT‐15 Flow Monitor
Operation and Programming Manual
16
During normal operation with the cover secured, you can change the Display Mode by touching the side
of the magnet wand to the MODE/DN location (top) on the outside of the housing. Use the MODE/DN
button to change the Display Mode when the RT‐15 cover is open. The RT‐15 displays the last mode
selected on power up. From seven of the display modes, you can also program the related variables.
LOGO Mode ‐ Turn On/Off Displays, or Change Modes
Use LOGO – Display Mode “RT‐15” during unit programming/set up to turn Display Modes on or off.
Turn off any of them based on your needs. To reassign, turn on, or turn off a display mode, use the
MODE/DN button (or magnetic switch) to select the LOGO display mode. Press and hold the ENT button
or activate and maintain the ENT magnetic switch. First a “P” appears in the lower left‐hand corner, and
then DISPLAY 00, which indicates which display mode is assigned to DISPLAY 00. DISPLAY 00 is assigned
to the LOGO screen allowing access for display activation/deactivation.
Turn off displays one at a time and return to LOGO display after each screen is off. Always turn off the
highest numbered displays first.
Turn Off Display Mode
To turn‐off a display mode, use the ENT button (or magnetic switch) to scroll through the DISPLAY #s
until you encounter the appropriate display mode. Use the RESET/UP button (or magnetic switch) to
turn the display off. The DISPLAY # assignment will indicate OFF DISPL. Use the ENT button (or magnetic
switch) to store the current assignment and advance the RT‐15 to the next available display number (00‐
09) or exit the LOGO display programming procedure.
Change Mode
To change a DISPLAY # assignment to a different mode, use the SEL button (or magnetic switch) to select
from the available display modes. Any DISPLAY # can be assigned as any display mode. Use the ENT
button (or magnetic switch) to store the current assignment and advance the RT‐15 to the next available
display number (00‐09), or exit the LOGO display programming procedure.
Turn Display Mode Back On
To restore a Display Mode that has been turned off, use the ENT button (or magnetic switch) to select a
DISPLAY # with an OFF DSPL assignment. Select a display mode assignment using the SEL button (or
magnetic switch). Use the ENT button (or magnetic switch) to store the current assignment and advance
the RT‐15 to the next available display number (00‐09), or exit the LOGO display programming
procedure.
Notice