Aquametrix P/R60C-6 Differential pH/ORP Probes User Manual
Page 2
![background image](/manuals/366265/2/background.png)
Model P60C-6 pH or R60C-6 ORP Differential Probes
www.WaterAnalytics.net
978-749-9949 Toll free - 855-747-7623 (855-pHProbe)
Technical Data
Related Products
Measuring Range
pH
0 to 14 pH
(Consult factory for applications below 2 or
above 12).
ORP
-1000 to +1000 mV
Flow Rate
10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in low
conductivity water and in solutions with
high suspended solids
Wetted Materials
CPVC, kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe)
Transmission Distance
3000 ft. (900 m)
Sensitivity
pH
0.01 pH
ORP
0.1 mV
Stability
pH
0.03 pH per day, non-cumulative
ORP
2 mV per day, non-cumulative
Temperature Copensation
Automatic
-5 to 95°C (23 to 203°F)
Pressure Limit
100 psig at 65°C maximum
Temperature Limits
CPVC
-5 to 95°C (23 to 203°F)
The temperature limit of probes in flow-
through applications is limited by pressure
and by the pipe fitting material.
Probe Cable
5 Conductor plus shield, 15 ft. (4.6m) long
CABLES & ACCESORIES
JB-1
NEMA 4X junction box
STC60-6
Mounting kit for submersion applications includes 1-1/4” NPT x 1” reducer, 4
feet of 1” CPVC pipe with water-tight strain relief fitting and securing assembly
P60HTC
Hot Tap Ball Valve assembly (Probe not included)
C42-5PXXX
Interconnect cable, dressed both ends. Specify length.
C35-17
Salt bridge Kit for all -6 series probes, package of 3.
C35-79
Compression fitting for all -6 series probes.
C34-51
Adapter for GLI Installations.
Protector-6
Protection shroud for submersion applications.
CALIBRATION SOLUTIONS
A35-13
pH 4 Buffer, 500 mL.
A35-14
pH 7 Buffer, 500 mL
A35-24
pH 10 Buffer, 500 mL
A35-40
ORP Buffer, 200 mV, 500 mL
A35-41
ORP Buffer, 600 mV, 500 mL
Ordering Information
P60C-6
pH probe complete with universal installation
fitting for flow-through or submersion applica-
tions
P60C-7
pH Probe for use with Hot-Tap hardware
P60C-6F
Flat-face pH probe
R60C-6
ORP probe complete with installation fitting
for low-through or submersion applications
R60C-6-G
ORP probe with gold electrode
R60C-7
ORP Probe for use with Hot-Tap hardware
Dimensions
18.1 (460)
14.0 (356)
1.125 (28.6)
1.00 (25.4)
9.25 (235)
5.08 (129)
1.125 (29)
1.00 (25.4)
P60C6