Model 2200c conductivity analyzer / controller, Technical data, Related products - probes & accessories – Aquametrix 2200C Conductivity Controller Datasheet User Manual
Page 2: Ordering information
![background image](/manuals/366262/2/background.png)
Model 2200C Conductivity Analyzer / Controller
www.WaterAnalytics.net
978-749-9949
Toll free - 855-747-7623 (855-pHProbe)
Technical Data
Related Products - Probes & Accessories
Display
4 x 7 segment 1/2” LED Display
Indicator Lights
4 LED indicators: Status, Alarm, Relay
A, Relay B
Measuring Range
Range
The range is factory set but may be field
changed
Temperature
0 to 100°C (32 to 212°F)
Power Requirements
Standard
98-132 VAC 50/60 Hz, (less than 10 VA)
Optional
196-264 VAC 50/60 Hz (less than 10 VA)
23-26 VDC (nominal current: 150 mA)
Ambient Conditions
Temperature
-30 to 50°C (-22 to 122°F)
Humidity
0 to 90% RH non-condensing
Control Relays
Number
Two Control Relays: Relay A, Relay B
Rating
5 A 115/230 VAC, 5A 30 VDC, SPDT
On/ Off Setpoints
Field selectable full scale
Cycle Timer
Field selectable 0 - 600 seconds
Manual Override
AUTO/OFF/ON
Fail-safe
Normal or fail safe operation
Alarm Relay
Number
One High and Low Alarm Relay
Rating
5 A, 115/230 VAC, 5A 30 VDC, SPDT
Alarm High
Field selectable full scale
Alarm Low
Field selectable full scale
Deadband
Fixed at 2% of full scale
Fail Safe
Normal or fail safe operation
Relay Indicators
Two individual LEDs indicate status
of two control relays and alarm relay
Analog Outputs
Non-isolated
0-1 mA, 100 ohms maximum load
Non-Isolated
0-5 VDC, 1000 ohms minimum load
Isolated
4-20 mA, 800 ohms maximum load
Output is isolated from the input, the
ground, line power and all other outputs.
Range Expand
Analog outputs can be scaled to rep-
resent any segment of the instrument
scale, minimum of 10% of full scale.
Output Hold
Analog outputs are placed on hold
when the instrument is in menu mode.
Temperature
The 0-1 mA and 0-5 VDC outputs
can be selected to track either the pH
or the temperature of the process.
Automatic
0-100°C (32 to 212°F)
Calibration
1 point, manual
Temperature Compensation
Automatic
0-100°C (32 to 212°F) if probe has tem-
perature sensor.
Diagnostics
Invalid Entries
Identified by flashing LEDs
Fail LED
Illuminates red to indicate Status Fault. Scroll
to Status in Menu to determine fault code. Re-
fer to Manual for probable cause and remedy
Auto LED
Illuminates red when Alarm Relay is energized,
indicating high/low process value or memory
loss.
Test
Display value and analog outputs can be set to
any value for testing and diagnostic purposes
Safety & Security
Operator
Password protected (activated by DIP switch)
Memory
Nonvolatile memory (EPROM)
Microprocessor
Watchdog timer monitors microprocessor. In-
strument automatically returns to online opera-
tion if left in menu mode for more than 10 min-
utes and no key is pressed.
Sensitivity
0.1% of span
Stability
0.1% of span per 24 hours non-cumulative
Temperature Drift
Zero - 0.01% of span per °C
Span - 0.01% of span per °C
Enclosure
NEMA 4X molded fiberglass reinforced polyes-
ter enclosure with four 1/2” conduit holes and
mounting feet for surface mount. A NEMA 4X
plug is provided for one hole.
Mounting
Standard
Surface Mount
Optional
Panel Mount kit P/N C35-68
Pipe Mount kit P/N C35-69
Weight
3.5 lbs. (1.6 kg)
2200C
Conductivity Analyzer/Controller
1
120 Vac, 50/60 Hz Power
2
240 Vac, 50/60 Hz Power
A
Surface Mount
B
Panel Mount
C
Pipe Mount
2200C
Ordering Information
MS Cells
Epoxy, low/high pressure, compression fitting 3/4” NPT.
MHMSC
Tank Mount Hardware for MS Cells.
MS-HTC
CPVC Wet/Hot Tap Mount for MS Cells.
Conductivity Range
(µSiemens/cm)
Cell
Constant
Conductivity Range
(mSiemens/cm)
Cell
Constant
0-2
0.01
0-10
10
0-5
0.01
0-20
10
0-10
0.05
0-50
10
0-20
0.05
0-100
20
0-50
0.05
0-500
50
0-100
0.05
0-1000
50
0-200
0.5
0-500
0.5
Special Ranges
0-1000
0.5
0-15% H
2
SO
4
50
0-2000
1
0-15% NAOH
50 tt
0-5000
1