Model 2200p ph analyzer / controller, Technical data, Related products - probes & accessories – Aquametrix 2200P pH Analyzer/Controller Datasheet User Manual
Page 2: Ordering information

Model 2200P pH Analyzer / Controller
www.Aquametrix.com
978-749-9949
Toll free - 855-747-7623 (855-pHProbe)
Technical Data
Related Products - Probes & Accessories
Display
4 x 7 segment 1/2” LED Display
Indicator Lights
4 LED indicators: Status, Alarm, Relay A,
Relay B
Measuring Range
pH
0.00 to 14.00 pH
Temperature
0 to 40
0
C (32 to 104
0
F)
Power Requirements
Standard
98-132 VAC 50/60 Hz, (less than 10 VA)
Optional
196-264 VAC 50/60 Hz (less than 10 VA)
23-26 VDC (nominal current: 150 mA)
Ambient Conditions
Temperature
-30 to 50°C (-22 to 122°F)
Humidity
0 to 90% RH non-condensing
Control Relays
Number
Two Control Relays: Relay A, Relay B
Rating
5 A 115/230 VAC, 5A 30 VDC, SPDT
Bidirectional
Relays factory preset for bidirectional
control, can be individually field set for ris-
ing or falling process
On/ Off Setpoints
Field selectable full scale
Fail-safe
Normal or fail safe operation
Alarm Relay
Number
One Alarm Relay
Rating
5A 115/230 VAC, 5A 30 VDC, SPDT
Alarm High
Field selectable 0 -14 pH
Alarm Low
Field selectable 0 -14 pH
Deadband
Fixed at 2% of full scale
Fail Safe
Normal or fail safe operation
Relay Indicators
Three individual LEDs indicate status
of two control relays and alarm relay
Analog Outputs
Non-isolated
0-1 mA, 100 ohms maximum load
Non-Isolated
0-5 VDC, 1000 ohms minimum load
Isolated
4-20 mA, 800 ohms maximum load
Output is isolated from the input, the
ground, line power and all other outputs.
Range Expand
Analog outputs can be scaled to rep-
resent any segment of the instrument
scale, minimum of 10% of full scale.
Output Hold
Analog outputs are placed on hold
when the instrument is in menu mode.
Temperature
The 0-1 mA and 0-5 VDC outputs can
be selected to track either the pH or
the temperature of the process.
Sensor to Controller Distance
Differential
P60 Series Differential pH Probe: maximum
distance 3000 feet (914 m)
Combination
500 Series: maximum distance 25 feet (7.62 m)
Calibration
2 point, manual or automatic buffer recognition
Temperature Compensation
Automatic
0-100°C (32 to 212°F) if probe has temperature
sensor.
Diagnostics
Invalid Entries
Identified by flashing LEDs
System Error
Two color STATUS LED indicates system sta-
tus. Green no fault, Red system fault. Scroll to
Status in Menu to determine fault code. Refer
to Manual for probable cause and remedy.
Alarm Relay
Alarm relay can be enabled to activate upon
high and low process values and memory loss.
Test
Display value and analog outputs can be set to
any value for testing and diagnostic purposes.
Safety & Security
Operator
Password protected (activated by DIP switch)
Memory
Nonvolatile memory (EPROM)
Microprocessor
Watchdog timer monitors microprocessor. In-
strument automatically returns to online opera-
tion if left in menu mode for more than 10 min-
utes and no key is pressed.
Sensitivity
0.1% of span
Stability
0.1% of span per 24 hours non-cumulative
Temperature Drift
Zero - 0.01% of span per °C
Span - 0.01% of span per °C
Enclosure
NEMA 4X molded fiberglass reinforced polyes-
ter enclosure with four 1/2” conduit holes and
mounting feet for surface mount. A NEMA 4X
plug is provided for one hole.
Mounting
Standard
Surface Mount
Optional
Panel Mount kit P/N C35-68
Pipe Mount kit P/N C35-69
Weight
3.5 lbs. (1.6 kg)
P60C-6
Industrial quality differential pH probe with replace-
able salt bridge, temperature sensor and integral
preamplifier. Compression fitting 1-1/4” NPT.
P60C-8
Industrial quality differential pH probe with replace-
able salt bridge, temperature sensor and inte-
gral preamplifier. Fixed threads 1-1/2” NPT.
P575
Industrial quality combination style pH
probe. Fixed threads 1” NPT.
P60-HTC
CPVC Wet/Hot tap ball valve assembly
STC60
Tank mount hardware for pH/ORP differential probes
2200P
pH Analyzer/Controller
1
120 VAC, 50/60 Hz Power
2
240 VAC, 50/60 Hz Power
A
Surface Mount
B
Panel Mount
C
Pipe Mount
2200P
Ordering Information