SHARK pH, ORP, Conductivity, Flow Controller / Analyzer

Technical Data

pH

ORP

Conductivity

Flow

Display

Front Panel: 4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED

Bar Graph. Inside Panel: 2 x 16 alpha-numeric LCD display

Measuring Range

pH: 0.01 to 14.00

Temp: 0 to 100°C

or 32° to +212°F

ORP: -1999 to +1999mV

(Dependent on sensor)

Temp: 0 to 100°C

or 32° to +212°F

MΩ/

cm

3

0 to 19.9

0.01

µS/

cm

3

0 to 2.00

0.01

0 to 20.0

0.1

0 to 200

0.1

0 to

2000

1.0

mS/

cm

3

0 to 20

10

0 to 200

50

Flow: 0 to 9999 with

selectable flow rate units

Volume: 0 to 9999 with

Auto Range Flow rate

units: Gallons (GP),

Cubic Feet (CF), Liters

(LP), Cubic Meters (CM),

custom by entering factor

related to Gallons.

Time units: Seconds (S),

Minutes (M), Hours (H)

Temperature

Compensation

Automatic or Manual

0 to 100°C (32° to +212°F)

Not required

Automatic or Manual

User selectable tem-

perature compensation

slope 0.0 to 10.0%/°C.

0 to 100°C (32° to +212°F)

Not required

Temperautre Unit

°C or °F

Not required

Temperature Sensor

User selectable: 300 Ω NTC Thermistor, 3000 Ω NTC Thermistor or Pt. 1000 RTD

Calibration modes

Auto-Calibration,

Manual Calibration,

Temperature Display

Manual Calibration

Temperature Calibration

Dry Calibration

Sample Calibration

Temperature Calibration

K factor input

Ambient Conditions

Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)

Menu Access

Front Panel

Auto-Calibration,

Manual Calibration,

Temperature Display

Manual-Calibration

Temperature Display

Manual Calibration

Temperature Display

Not available

Menu Access

Inside Panel

Full Access to all parameters of operations menu

Relay Outputs

Two Control Relays, 10 A / NO, 5 A / NC @ 240 VAC or 28 VDC.

Mode: Process control, Adjustable parameters: process direction,(rising or falling) on-set-point, off set-point,

(0 to 100% of full scale), cycle timer (on / off, 0 to 600 seconds), failsafe (on / off).

One Alarm Relay, 10 A / NO, 5 A / NC @ 240 VAC or 28 VDC.

Mode: High / Low Alarm, Adjustable parameters: Low on / Low off set-

point (0 to 100% of full scale, low on must be less than

low off), High On / High Off set-point (0 to 100% of full scale, High on must be greater than High off).

Analog Outputs

4 to 20 mA Channel 1

Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800 Ω

4 to 20 mA Channel 2

Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800 Ω

Can be set to track temperature if sensor is equipped with a temperature sensor

Memory Back-up

All user settings are retained indefinitely in memory (EEPROM)

Mechanical

Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with four 1/2” conduit holes

Mounting: Universal Mounting kit for surface, pipe and panel mount, is included

Sensor Input

Probe: -600 to +600 mV

Temp. Sensor: 0 to 9999 Ω

Probe: -1999 to

+1999 mV

Temp. Sensor:

0 to 9999 Ω

Cell: 0 to 9999 Ω

Temp. Sensor: 0 to 9999 Ω

Paddle: 0 to 2000 Hz

Invalid Entries

Invalid entries cannot be stored

Manual Test Mode

Process value can be simulated with arrow keys to verify correct setup of outputs

Manual

Relay Override

Relays can be set to on / off / auto, to verify correct wiring of auxiliary devices, or to manually adjust process

Auto Return

User selectable auto return if SHARK is left in menu mode or if re-

lays are left in manual override mode for more than 10 min.

Display Damping

User can select rate at which SHARK updates display. Enables display damping of unstable process

Net Weight

2.2 lbs (1 kg)