Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the

FUNC key and keep it de-

pressed for 2 seconds.
FIL and the current setting appear in

the display:
0 = slight damping

⇒ fast determination of weight

1 = average damping

⇒ normal determination of weight

2 = strong damping

⇒ slow determination of weight

– Press an arrow key if you want to

change the set value.

– Press the

FUNC key again to store the

displayed value. The scale is then back

in normal weighing mode.

The settings are retained even once the scale is switched off.

6. Cleaning

Clean the rubber surface and the housing as required using a domestic cleaning agent or

commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for

cleaning. Such substances can damage the high-quality surfaces.

7. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Is the plug contact of the mains unit properly connected to the scale?
– Is the plug contact for connecting the display to the scale correctly connected?

0.0

does not appear before weighing?

– Press the Start key again – there must not be any load on the scale – and only its feet

should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.