Store weight value (HOLD)

You can also continue displaying the weight determined once the load has been removed.

This makes it possible to look after a patient first, before having to note the weight:
– Press the

HOLD/TARE key whilst the

load is on the scale.

“HOLD” appears in the digital display.

The value is stored until the scale

switches off automatically.
The triangle with the exclamation mark

indicates that this frozen display is an

additional facility which is not subject to

calibration.

– The HOLD function can be switched off

by pressing the

HOLD key again.

0.0

appears in the display if the scale

has no load on it, otherwise the current

weight. The “HOLD” display vanishes.

Note:

The HOLD function can also be used if the TARE function has been executed

beforehand.

Body-Mass-Index (BMI)

Body Mass Index puts height and weight in relationship to one another, providing more
accurate information than ideal Broca weight, for example. A tolerance range is given
which is considered ideal for health.
– Press the Start key with no load on the

scale.

se(A

,

8.8.8.8.8.

and

0.0

appear conse-

cutively in the display. The scale is then

automatically set to zero and ready for

operation.

– Press the

FUNC key.The last function

used appears in the display.

– Use the arrow keys to select the BMI

function

FU 1.

– Activate the function by pressing the

FUNC key once again.
The patient’s recorded height will then
appear on the display.

– You can use the arrow keys to adjust

the value in increments according to the

graduations displayed.

– Once the correct value is set, press the

FUNC key again. The BMI function is

now activated.