Troubleshooting

DANGER

There is a danger of injury and possible equipment damage through in

correct power supply!Let problems relating to an incorrect power supply

be eliminated by an electrical specialist! Do not carry out arbitrary work

on the electrical components of the pyrometer!

Only eliminate such problems yourself when their causes obviously relate to incorrect
power supply, under-cooling or contamination of the lens.

Do not undertake any interventions into the pyrometer. If problems arise which do not
relate to the causes mentioned above, inform the service staff of the
KLEIBER Infrared GmbH (for contact data see page 22).

Fault

Cause

Solution

Pyrometer does not provide
any measured values

Power supply faulty or
interrupted

- Check the power supply
- Check plugs and connections
- Inspect cable

If the pyrometer supplies
inaccurate measured values
or measured values which
lie outside the range to be
expected

dirty optics or condensation
on the lens

Clean optics (see section 9.2)