8troubleshooting – LumaSense Technologies IGA 740-LO User Manual
Page 24
20
8
Troubleshooting
DANGER
There is a danger of injury and possible equipment damage through in
correct power supply!Let problems relating to an incorrect power supply
be eliminated by an electrical specialist! Do not carry out arbitrary work
on the electrical components of the pyrometer!
Only eliminate such problems yourself when their causes obviously relate to incorrect
power supply, under-cooling or contamination of the lens.
Do not undertake any interventions into the pyrometer. If problems arise which do not
relate to the causes mentioned above, inform the service staff of the
KLEIBER Infrared GmbH (for contact data see page 22).
Fault
Cause
Solution
Pyrometer does not provide
any measured values
Power supply faulty or
interrupted
- Check the power supply
- Check plugs and connections
- Inspect cable
If the pyrometer supplies
inaccurate measured values
or measured values which
lie outside the range to be
expected
dirty optics or condensation
on the lens
Clean optics (see section 9.2)