16

Pyrometer wire connection

6.2.2 Align pyrometer

The pyrometer is equipped with an LED pilot light for accurate alignment of

the sensor with the object to be measured. Align the pyrometer with the object

to be measured as follows:

1.

Switch on the supply voltage to the pyrometer.

2.

Allow a starting time of approx. 5 minutes for thermo-stabilization. Than

the pyrometer has stabilised and is ready to work with the given

accuracy.

3.

Switch on the pilot light by pressing the push button at the rear side.

Make sure that there is nothing in the path of rays.
Note: In order to avoid measuring errors, the area of the pilot light must

not be larger than the object to be measured

4.

If you have a pyrometer with vario optic head, adjust the necessary

measuring distance (see page 18).
Note:

The pyrometer is thus aligned and ready for the temperature

measurement.

1

Supply voltage +24 V (DC or AC)

2

Supply voltage 0 V (DC or AC)

3

ground output

4

Analogue output [0 (4) … 20] mA

5

shiled

1 - white

Supply voltage +24 V (DC or AC)

2 - brown

Supply voltage 0 V (DC or AC)

3 - green

ground output

4 - yellow

Analogue output [0 (4) … 20] mA

5 - green/yellow

shield