Finishing the settings 10 – Yokogawa µR20000 User Manual
Page 42
42
IM 04P02B01-02E
Finishing the Settings
10.
When
Setting complete screen is displayed, do either of the following:
Press the key to set other alarms.
To finish setting the alarm, press the
ESC key.
02-02 CH/level 1
Setting complete
11.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Explanation
In step 7, you can select an alarm type on the table below.
Type
Description
H
High Limit Alarm: An alarm occurs when the input value exceeds the alarm value.
L
Low Limit Alarm: An alarm occurs when the input value falls below the alarm value.
h
Difference High Limit Alarm
*1
: An alarm occurs when the difference in the input values
of two channels is greater than or equal to the specified value.
l
Difference Low Limit Alarm
*1
: An alarm occurs when the difference in the input values
of two channels is less than or equal to the specified value.
R
High Limit on Rate-of-Change Alarm
*2
: The rate-of-change of the measured values
is checked over a certain time (interval). An alarm occurs if the rate-of-change of the
measured value in the rising direction is greater than or equal to the specified value.
r
Low Limit on Rate-of-Change Alarm
*2
: The rate-of-change of the measured values is
checked over a certain time (interval). An alarm occurs if the rate-of-change of the
measured value in the falling direction is greater than or equal to the specified value.
T
Delay High Limit Alarm
*3
: An alarm occurs when the measured value remains above
the alarm value for a specified time period (alarm delay period).
t
Delay Low Limit Alarm
*3
: An alarm occurs when the measured value remains below
the alarm value for a specified time period (alarm delay period).
*1 Can be specified on channels set to delta computation.
*2 You must set the interval in Basic Setting mode.
*3 You can select T or t when the alarm delay function is enabled in Basic Setting mode.
Setting the Input Range and Alarm on Measurement Channels