Yokogawa µR20000 User Manual
Page 39
39
IM 04P02B01-02E
Selecting the Input Type
5.
Press the
key to select
Scale, and press the key (for the selectable
settings, see “Explanation” on page 40).
Mode=Scale
Scales and records the
6.
Press the
key to select
Volt and then press the key.
Type=Volt
DC Voltage
7.
Press the
key to select
20V and then press the key.
Range=20V
20mV-50V
Setting Span Left
8.
Set Span left to
0.00, and press the key.
Press the
key to select the desired digit. Press the
key to select the
value.
Span left= 0.00
-20.00/ 20.00V
Displays the range of span left.
Setting Span Right
9.
Likewise, set Span right to
10.00, and press the key.
Span right= 10.00
-20.00/ 20.00V
Displays the range of span right.
Setting the Decimal Position and Scale Left
10.
Display Scale left.
Scale left= 0.00
-200.00/ 300.00
Displays the scaling range.
11.
Press the
key to select the desired digit.
Scale left= 0.00
-200.00/ 300.00
12.
Press the
key to select space and then press the key (Scale left is set to
0.0).
Scale left= 0.0
-200.00/ 300.00
Setting Scale Right
13.
Likewise, set Scale right to
400.0, and press the key.
Scale right= 400.0
-2000.0/ 3000.0
The
Setting complete screen is displayed. When this screen is displayed, the settings
entered up to then are applied.
Finishing the Settings
14.
When the
Setting complete screen is displayed, press the ESC key.
The
Set=Range screen is displayed.
04-04 Channel
Setting complete
Set=Range
Input range and record
Setting the Unit and Finishing the Unit Settings
See steps 13 to 17 in Setup Example (2).
15.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Setting the Input Range and Alarm on Measurement Channels
* When the key or key is pressed while holding down the
SHIFT key, the operation is reversed as
when the respective key is pressed by itself.