Yokogawa µR20000 User Manual

Page 39

background image

39

IM 04P02B01-02E

Selecting the Input Type

5.

Press the

key to select

Scale, and press the key (for the selectable

settings, see “Explanation” on page 40).

Mode=Scale

Scales and records the

6.

Press the

key to select

Volt and then press the key.

Type=Volt

DC Voltage

7.

Press the

key to select

20V and then press the key.

Range=20V

20mV-50V

Setting Span Left

8.

Set Span left to

0.00, and press the key.

Press the

key to select the desired digit. Press the

key to select the

value.

Span left= 0.00

-20.00/ 20.00V

Displays the range of span left.

Setting Span Right

9.

Likewise, set Span right to

10.00, and press the key.

Span right= 10.00

-20.00/ 20.00V

Displays the range of span right.

Setting the Decimal Position and Scale Left

10.

Display Scale left.

Scale left= 0.00

-200.00/ 300.00

Displays the scaling range.

11.

Press the

key to select the desired digit.

Scale left= 0.00

-200.00/ 300.00

12.

Press the

key to select space and then press the key (Scale left is set to

0.0).

Scale left= 0.0

-200.00/ 300.00

Setting Scale Right

13.

Likewise, set Scale right to

400.0, and press the key.

Scale right= 400.0

-2000.0/ 3000.0

The

Setting complete screen is displayed. When this screen is displayed, the settings

entered up to then are applied.

Finishing the Settings

14.

When the

Setting complete screen is displayed, press the ESC key.

The

Set=Range screen is displayed.

04-04 Channel

Setting complete

Set=Range

Input range and record

Setting the Unit and Finishing the Unit Settings

See steps 13 to 17 in Setup Example (2).

15.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Setting the Input Range and Alarm on Measurement Channels

* When the key or key is pressed while holding down the

SHIFT key, the operation is reversed as

when the respective key is pressed by itself.