1. Outline
1.2 Personal Conductivity Meter Specifications
Applicable measurement range:
• Sensors for high purity water measurement (cell constant: 0.05 cm
-1
)
Conductivity:
0 to 2
S/cm, 0 to 20S/cm, 0 to 200S/cm, 0 to 40M⍀•cm
Temperature:
0 to 80
؇C*
1
• For general-purpose sensors (cell constant: 5 cm
-1
)
Conductivity:
0 to 20
S/cm, 0 to 200S/cm, 0 to 2mS/cm, 0 to 20mS/cm,
0 to 200mS/cm,
Temperature:
0 to 80
؇C*
1
• For chemical-corrosion-resistant sensors (cell constant: 5 cm
-1
)
Conductivity:
0 to 20
S/cm, 0 to 200S/cm, 0 to 2mS/cm, 0 to 20mS/cm,
0 to 200mS/cm
Temperature:
0 to 80
؇C*
1
• Sensors for high-conductivity measurement (cell constant: 50 cm
-1
)
Conductivity:
0 to 2mS/cm, 0 to 20mS/cm, 0 to 200mS/cm, 0 to 2S/cm
Temperature:
0 to 80
؇C*
1
Resolution: Conductivity:
Least significant digit: 1 digit;
Temperature:
0.1
؇C
Repeatability (Conductivity):
؎2% (؎5% for general-purpose sensor on 0 to 200mS/cm measurement range)
Accuracy (Temperature):
؎0.7؇C (in the range 0 to 70؇C)
؎1؇C (when exceeding 70؇C)
Temperature compensation range:
variable 0.00 to 9.99%/
؇C, also (preset) NaCl solution temperature coefficient
Measurement converted to conductivity at standard temperature of 25
؇C
Measured liquid temperature: 0 to 80
؇C (0 to 50؇C when the sensor cable is immersed in
water)
Ambient temperature:0 to 50
؇C
Construction:
IEC 60529, IP67 class of protection from environment
Power requirement:
Two of size AA batteries (LR6)
Automatic power off interval: May be set in range 1 to 120 minutes
Battery life:
About 200 hours of continuous use, for long-life alkaline battery (life
may vary depending on battery type and ambient conditions)
External dimensions: Approximately 150 (H) x 61 (W) x 42 (D) mm (not including
protruding portions)
Weight:
Approximately 220 g (including batteries)
*1: Display range is from -10 to 120
؇C.