IM 12D03D02-01E

5-6

5. Handling of the SC72 Personal Conductivity Meter

Procedures for calibration with standard solution

F050401.EPS

Whole measurement value
is flashing

Complete the calibration

Least significant figit of the
measuremet value is flashing

Select
CAL mode

POWER

F/ENT

F/ENT

F/ENT

F/ENT

Std. sol.

Water

Wash sensor with
water completely.

Wait until the measured
value is stable.

Set the conductivity of
the standard solution.

Thoroughly wipe off the
washing water from the
electrodes before immersing
the sensor in standard solution.

Refer to Section 4.3, “Function Modes,” for temperature coefficient and cell constant
settings.