5. Handling of the SC72 Personal Conductivity Meter

5. Handling of the SC72 Personal

Conductivity Meter

5.1 Tips to Maintain Meter Performance

The SC72 meter appears to be very simple, but is a precision instrument. To ensure that
measurement accuracy is maintained, observe the following precautions regarding
preliminary setting, measurement, maintenance and storage.

Preliminary setting, measurement, maintenance and storage:

Measurement

Maintenance

Preliminary

setting

Storage

F050101.EPS

Table 5.1 Observe the following precautions

P r e l i m i n a r y
s e t t i n g

M e a s u r e m e n t

M a i n t e n e n a c e

S t o r ag e

T0501.EPS

Setting cell constant

• Set specific electrode cell constant.

Setting temperature coefficient

• Set liquid temperature coefficient when the standard temperature

conversion is required (see Section 2.5)

If standard temperature conversion is not requiresd, set 0.00.

• Select one of the three conductivity measurement ranges (0 to

200

S/cm, 0 to 200mS/cm, or 0 to 2S/cm) corresponding to the type of

electrode used.

Temperature measuring range is 0 to 80

؇

C.

• When measurement is completed, thoughly wash electrode to remove the

measured liquid.

• If the electrode gets stained or deformed, its cell constant may change.

In this case, calibrate the meter with standard solution (see Section 5.4).

• Avoid high-temperature, high-humidity storage locations.