Yokogawa SC72 Personal Handheld Conductivity Meter User Manual

Page 25

background image

IM 12D03D02-01E

3-4

3. Measurement

(2) Data storage

If the

DATA

key is pressed during measurement,

mark starts flashing. Press the

F/ENT

key, then currently measured data can be stored in nonvolatile memory. Data

stored are measured conductivity, measured temperature, date and time. Up to 300 data
including individually deleted data can be stored. If you attempt to store more data,

is displayed.

If

is displayed even though the number of data stored is less than 300, perform

“defrag” [refer to Section 4.3 (15), “Defragment storaged data (DFLG) panel”]. By doing
so, you can increase the number of data to be stored.

To check stored past data, refer to 4.3 (1) dAt (display stored measurement value) later
in this manual.

While

is flashing, press the

DATA

key or

MEAS

key to return to measurement

mode without storing data.

F030402.EPS

Abort data storage.

Execute data storage.

Press key to store data while is flashing.

While is flashing, press
or keys to abort data storage.

F/ENT

MEAS

DATA

DATA