4 setting temperature unit, 5 setting cell constant – Yokogawa SC72 Personal Handheld Conductivity Meter User Manual
Page 17

IM 12D03D02-01E
2-4
2. Preparation
2.4 Setting Temperature Unit
Default temperature units are
؇C. To change to ؇F, refer to Sec. 4.3 (13) Set temperature
units (tP.U) panel.
2.5 Setting Cell Constant
Even for sensors of the same type, each sensor has its own distinct cell constant. So, set
the proper cell constant as indicated on the sensor cable.
Whenever sensors are replaced, be sure to change the cell constant setting in the meter
accordingly. Cell constants once set are stored in non-volatile memory and are not lost
even when the batteries are replaced.
• Setting cell constants
Press the
F/ENT
key to switch to function mode. Then select the C.C display with the
key and access the cell constant setting display with the
F/ENT
key. Use the
key to set the cell constant, then press the
F/ENT
key to confirm it.
Refer to Sec. 4.3 (4).
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)