6. Troubleshooting

6.2 Error Messages and Corrective Action

Table 6.1

Error Message

T0601.EPS

Error message

Err1

Err2

Err3

Err6

or

Name

Temperature compensation computation value out of range

Out of temperature measurement range

Calibration error

Meter electronics failure

Out of measuring range

*1 Alphabet values simulated on a numeric display

*1

(1) Err1

Temperature compensation computation value out of range

Cause:

When the standard temperature conversion is made to conductivity at 25

؇C, this

instrument performs temperature compensation computation based on the preset
temperature coefficient.

Err1 is displayed in cases where temperature compensation computation is impossible
because the preset temperature coefficient is improper or a solution with a large
temperature coefficient is measured at temperatures lower or higher than 25

؇C.

Corrective action:

* Reset the temperature coefficient to the correct value.

* Measure solution at about 25

؇C.

(2) Err2

Out of temperature measurement range

Cause:

This error occurs if the measurement liquid temperature is outside the range -10 to
120

؇C, or if sensor cables are disconnected.

Note: The liquid temperature should never be outside the range 0 to 80

؇C.

Corrective action:

* Set the liquid temperature to 0 to 80

؇C.

* Replace the sensor. (Refer to Sec. 1.6)