2 error messages and corrective action, 2 error messages and corrective action -2 – Yokogawa SC72 Personal Handheld Conductivity Meter User Manual
Page 53

IM 12D03D02-01E
6-2
6. Troubleshooting
6.2 Error Messages and Corrective Action
Table 6.1
Error Message
T0601.EPS
Error message
Err1
Err2
Err3
Err6
or
Name
Temperature compensation computation value out of range
Out of temperature measurement range
Calibration error
Meter electronics failure
Out of measuring range
*1 Alphabet values simulated on a numeric display
*1
(1) Err1
Temperature compensation computation value out of range
Cause:
When the standard temperature conversion is made to conductivity at 25
؇C, this
instrument performs temperature compensation computation based on the preset
temperature coefficient.
Err1 is displayed in cases where temperature compensation computation is impossible
because the preset temperature coefficient is improper or a solution with a large
temperature coefficient is measured at temperatures lower or higher than 25
؇C.
Corrective action:
* Reset the temperature coefficient to the correct value.
* Measure solution at about 25
؇C.
(2) Err2
Out of temperature measurement range
Cause:
This error occurs if the measurement liquid temperature is outside the range -10 to
120
؇C, or if sensor cables are disconnected.
Note: The liquid temperature should never be outside the range 0 to 80
؇C.
Corrective action:
* Set the liquid temperature to 0 to 80
؇C.
* Replace the sensor. (Refer to Sec. 1.6)