2 principles of operation, 2 principles of operation -2 – Yokogawa SC72 Personal Handheld Conductivity Meter User Manual
Page 57

IM 12D03D02-01E
7-2
7. Measuring Principles of this Instrument
7.2 Principles of Operation
F0702.EPS
AC power supply
Ammeter
Figure 7.2
Operational Schematic
Dip two metallic plates (used as sensor electrodes) in a solution and apply a certain
voltage to flow current (I):
I =
(1)
E
Rc
where Rc = solution resistance between two sensors
The relationship between Rc and conductivity K is expressed by:
Rc = J •
(2)
1
K
where J = constant (also called cell constant) — determined by
sensor and distance between two sensors
From equations (1) and (2) above, we obtain:
K =
• I
(3)
J
E
E and J are constant, so the conductivity (K) can be obtained by
measuring the current (I).
See also other documents in the category Yokogawa Sensors:
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)