2 alarms and countermeasures, Alarms and countermeasures -3, Table 4.2 alarm message summary – Yokogawa EJX440A User Manual
Page 32
IM 01C25T03-01E
4-3
4. SELF-DIAGNOSTICS
4.2 Alarms and Countermeasures
Table 4.2 Alarm Message Summary
Indicator
BT200 display
Cause
Countermeasure
Output operation
during error
None
GOOD
AL. 01
CAP. ERR
01: P-SENSOR ERR
AL. 02
AMP. ERR
02: AT-SENSOR ERR
AL. 11
ST. PRSS
11: SP OVER SPEC
AL. 12
CAP. TMP
12: CT OVER SPEC
AL. 13
AMP. TMP
13: AT OVER SPEC
AL. 30
RANGE
30: P OVER RANGE
AL. 35
P. HI
35: P HIGH ALARM
AL. 36
P. LO
36: P LOW ALARM
AL. 37
SP. HI
37: SP HIGH ALARM
AL. 38
SP. LO
38: SP LOW ALARM
AL. 39
TMP. HI
39: CT HIGH ALARM
AL. 40
TMP. LO
40: CT LOW ALARM
AL. 51
P. URV
51: P ILLEG URV
AL. 52
P. SPN
52: P ILLEG SPAN
AL. 53
P. ADJ
53: P SPAN ADJ
53: P ZERO ADJ
AL. 54
SP. RNG
54: SP ILLEG LRV
54: SP ILLEG URV
54: SP ILLEG SPAN
AL. 55
SP. ADJ
55: SP SPAN ADJ
AL. 60
SC. CFG
60: SC CONFIG ERR
55: SP ZERO ADJ
Static pressure exceeds
limit.
Capsule temperature is outside
range (–50 to 130
°
C).
Amplifier temperature is outside
range (–50 to 95
°
C).
Output is outside upper or lower
range limit value.
AL. 31
SP. RNG
31: SP OVER RANGE Static pressure exceeds
specified range.
Input pressure exceeds specified
threshold.
Input static pressure exceeds
specified threshold.
Detected temperature exceeds
specified threshold.
AL. 50
P. LRV
50: P ILLEG LRV
Specified value is outside of
setting range.
Specified values or settings do
not meet the conditions.
T0401.EPS
Sensor problem.
01: CT-SENSOR ERR
Capsule temperature sensor
problem.
01: C-EEPROM ERR
Capsule EEPROM problem.
Amplifier temperature sensor
problem.
02: A-EEPROM ERR
Amplifier EEPROM problem.
02: CPU BOARD ERR Amplifier problem.
Continues to operate and
output.
Continues to operate and
output.
Continues to operate and
output.
Continues to operate and
output holding static
pressure in %.
Outputs the signal (Hi or Low)
set with parameter D26.
[Status output: undefined]
Use heat insulation or
make lagging to keep
temperature within
range.
Check input and range
setting, and change
them as needed.
Check input.
Adjust settings and
change them as needed.
Adjust settings and
change them as needed.
Check settings and change
them as needed.
Check settings and change
them as needed.
AL. 79
OV. DISP
Displayed value exceeds
limit.
Check settings and change
them as needed.
Replace capsule when
error keep appearing error
even after restart.
Replace capsule.
Replace amplifier.
AL. 10
PRESS
10: P OVER SPEC
Input is outside measurement
range limit of capsule.
Outputs upper range limit (URL)
or lower range limit (LRL).
Outputs upper range value
(URV) or lower range value
(LRV).
Continues to operate and
output.
Holds output immediately
before error occurred.
Check input or replace
capsule when necessary.
Check capsule
temperature.
Continues to operate and
output without signal
characterizing.
Continues to operate and
output.