Yokogawa EJX440A User Manual
Page 18
IM 01C25T03-01E
3-10
3. OPERATION
a. Display Selection
(I10: DISP OUT1)
Select the variable for the parameter I10: DISP OUT1
to display on the integral indicator.
• Example: Change the integral indicator scale
from % of range to input pressure
display.
ESC
SET
I10:DISP OUT1
PRES
FEED
NO
OK
Use the or key
to select PRES.
Press the key twice to
enter the setting.
Press the (OK) key.
F0318.EPS
SET
I10:DISP OUT1
PRES %
< PRES >
< PRES % >
< ENGR.PRES >
< SP >
b. Cyclic Display
(I11: DISP OUT2, I12: DISP OUT3, and I13:
DISP OUT4)
In addition to the display set at I10: DISP OUT1,
displays can be set at I11: DISP OUT2, I12: DISP
OUT3, and I13: DISP OUT4 for cyclic display in the
order of the parameter number.
c. User Setting of Engineering Unit and Scale
(I30: ENGR.UNIT, I31: EASY EU SET, I33:
ENGR.LRV, and I34: ENGR.URV)
These parameters allow the entry of the engineering
units and scale to be displayed. The engineering unit
can be selected from the parameter of I31: EASY EU
SET as listed below. Alternately, up to eight alphanu-
merics, spaces, and a slash “/” can be input on keypad
at I30: ENGR.UNIT; only first six are displayed on the
integral indicator.
Select the unit from the list of I31: EASY EU SET.
FX0302.EPS
kPa
MPa
mbar
bar
psi
psia
mmH2O
mmHg
mmHgA
mmAq
mmWG
Torr
inH2O
inHg
inHgA
ftH2O
gf/cm2
kgf/cm2
kg/cm2G
kg/cm2A
atm
kg/h
t/h
m3/h
m3/min
l/h
l/min
kl/h
kl/min
Nl/h
Nl/min
Nm3/h
Nm3/min
ACFH
ACFM
SCFH
SCFM
GPH
GPM
m
mm
in
ft
kg/m3
g/cm3
Follow the procedure below to change the settings.
• Example: Set an engineering unit M.
CODE
CAPS
CLR
ESC
SET
I30:ENGR.UNIT
M
FEED
NO
OK
SET
I30:ENGR.UNIT
M_
Set M.
Press the key twice to
enter the setting.
Press the (OK) key.
F0319.EPS
Note that following symbols are not available.
. – , + * ) ( ’ & % $ # ” !
The transmitter integral indicator shows “-- -- -- -- -- --”
when these are entered.
• Example: Set lower range value (LRV) to –50
and upper range value (URV) to 50.
FEED
NO
OK
SET
I34:ENGR.URV
50 M
Press the (OK) key.
F0320.EPS
DEL
CLR
ESC
DATA
DIAG
PRNT
ESC
PARAM
I32:ENGR.EXP
---
I33:ENGR.LRV
– 50 M
I34:ENGR.URV
50 M
DEL
CLR
ESC
SET
I33:ENGR.LRV
0.00 M
- 50
Set –50.
Press the key twice to
enter the setting.
Setting LRV
Setting URV
Set 50.
Press the key twice to
enter the setting.
SET
I34:ENGR.URV
100.00 M
+ 50
FEED
NO
OK
SET
I33:ENGR.LRV
- 50 M
Press the (OK) key.