HAPTER

4

L

IFE

C

YCLE

C

OUNTERS

78

9650-0801-01 Rev B

CO2

On the Propaq MD, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes.

3.

Select Reset CRC Counter. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.

SpO2

On the Propaq MD, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>SpO2>Error Counts.

3.

Select Reset CRC Counter. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.

Defibrillator/Pacer

On the Propaq MD, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>Defib/Pacer>Error
Counts
.

3.

Select Reset CRC Counter. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.