23.0 Temperature Test

Equipment

IBP Temperature Simulator (or equivalent)

Do this...

Observe This...

Pass/Fail

23.1 Connect the temperature simulator to the

Propaq MD T1 channel.

23.2 On the IBP Temp Simulator, Select 98.6

Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.

The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.

o

o

23.3 Connect the Temperature simulator to the

Propaq MD T2 channel

23.4 On the IBP Temp Simulator, Select 98.6

Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.

The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.

o

o

23.5 Disconnect cable.