beautypg.com

ZOLL Propaq MD Rev B User Manual

Page 52

background image

C

HAPTER

2

T

ROUBLESHOOTING

48

9650-0801-01 Rev B

IBP DEVICE FAILURE

IBP software detected a
problem with the IBP data
that is most likely a hardware
issue.

Replace the main board.

IBP PROBE1 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE2 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE3 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the IBP probe. If
problem persists, replace
the right side panel.

IBP PROBE1 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

IBP PROBE2 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

IBP PROBE3 CIRCUIT
FAILURE

An unsupported probe was
inserted or there is a circuit
problem.

Unplug probe, power cycle.
If problem persists, replace
main board.

SPO2 DEVICE FAILURE

Communication failure or
device malfunction.

Check SpO

2

communication cables. If
problem persists, replace
the SpO

2

module.

NIBP MONITORING
FAILURE

One of the following issues
occurred:

A failure was reported by
the NIBP module.

The two transducers are
not aligned. They are
either not calibrated,
there is a transducer
failure, or there is a
kinked tube.

A calibration problem was
reported by the NIBP
module.

If problem persists, replace
the NIBP module.

Error Message

Explanation

Technical Action