4. when replace the battery. The correct connection is : connect the red line

to the plugs in red mark; and the black line to the plugs in black mark, Be sure

wiring is correct, or it will cause damage to load cell, explosive, and the main

board damage

5. Our Company only responsible for approved the standard battery and

power adapter by our company. And not take any responsibility for any damage

by the accessories offered by third-party.

2 Main functions and parameter

2.1 main functions:

Weighing function:

Zero, gross, net, accumulating, printing, counting, animal -weighing.

Toggle operation: kg/lb

Options:

Pinter

RS232/RS485 serial interface or second display

Classifying scales (with I/O card)

2.2 technical parameter

Accuracy class

6000 e

Resolution

display: 30, 000 ADC: 2,000,000

Zero stability error TK

0

< 0.1

μV//K

Span stability error TK

spn

<

± 6 ppm//K

Conversion time

50 ms

Sensitivity (internal) 0.12

μV /count

Input voltage

-30~30mV DC

Excitation circuit 5 VDC, 4 wire connection, all strain gauge,

min. 50

Ω, e.g.: 6 load cell of 350Ω

AC power AC100~250V
DC power DC6V/4Ah rechargeable batter,700mA。

Operation temperature

- 10

°C ~ + 40 °C

Storage temperature

- 40

°C ~ + 70 °C