Vestil SCALE-S User Manual
Page 26
26
5. Error resolution and daily maintenance
Error code list
Error code
Reason
Resolution
UUUUUU
Overflow:
measuring value is above FSD
+ overload range
1.Take off the goods from
scales
2.Recalibration
3. Check load cell
4.Chang main board
nnnnnn
Underflow:
measuring value is below
negative display range
1.Recalibration
2. Check load cell
3.Chang main board
ERR1
During calibration:
no enter the calibration weight
value
Input
weight
of
the
calibrated weights
ERR2
During calibration:
the used calibration weight
value is too low
Add weights.
Recommend the weights
is 15-80% the Max.
capacity
ERR3
During calibration:
input voltage is negative
1. Check the installation is
ok or not
2.Check the connection
for load cell is ok or not
ERR4
During calibration:
measuring
value
is
not
standstill
Check the scales
installation is ok and the
make sure the goods on
the scales is stable
ERR5
Checksum error of EEPROM
1.Power off the indicator
and re open again
2. Change main board
Lobat
The voltage of rechargeable
battery is too low
Recharge the battery