Vestil SCALE-S User Manual
Page 21
21
can have the overload or under load alarming function. As follows
Menu
Optional function
Parameter
C 20
Value for upper limit
alarm on
When the weight reach this value.
Indicator output the overload signal
C 21
Value for upper limit
alarm off
When the weight reach this value.
Indicator stop the overload signal
C 22
Value for lower limit
alarm on
When the weight reach this value.
Indicator output the underload signal
C 23
Value for lower limit
alarm off
When the weight reach this value.
Indicator stop the underload signal
1. when set C20 C21 C22 C23 default=000000, it means close the
upper limit(overload) and lower limit(under load) alarm
2. when set C20=C21, C22=C23, it means open upper
limit(overload) and lower limit(under load) alarm. It mainly suit for
classifying mode. For example:
The accepted range for a bag of rice is 24.9-25.0kg, setting as follows
1). Press
and
. Till it show C01. enter into setting
parameter
2). Use
← → ↑ ↓ to set the C20. Press
. And it show
“0000
00
”
3). Follow the above steps. Set the C20 C21 C22 C23 as follows
Upper limit: C20=C21=25.10 kg
Lower limit: C22=C23=24.90 kg