Vestil SCALE-S User Manual

Page 24

background image

24

Menu

Option function

Parameter setting

C27

Communication

mode

C27=0 means shut off

C27=1 means continuous sending.

Connect to big display

C27=2 connect printer

C27=

3

command

mode

from

supervisory

C28

Serial interface

Baud rate

C28=0 baud rate=1200

C28=1 baud rate=2400

C28=2 baud rate =4800

C28=3 baud rate=9600

C29

Data bit

and calibration bit

C29=0 means 8,none

C29=1 means 7,even

C29=2 means 7,odd

2. Second display using method

Connect with second display according to the

“3.2.3” in this manual

and Set C27=1. and C28 & C29 the same data bit and calibration with

second display. After finish setting. Restart the indicator and second

display. If the same data shown on the indicator and second display.

Means it works regularly

3. Printer operation:

This indicator can connect with built-in mini serial interface printer

and outside serial interface portable printer. And built-in printer is setted

Within the factory

Indicated need setting according to the communication data for the

outside portable printer.

1. communication method set: Press

and

enter