ATO-400P / ATO-600P Operating Procedures

39

Table 13.0 Calibration Test Procedures (Continued)

STEP

ACTION

ATO DISPLAY

13-9

Automatic; the ATO will measure the 100

?

-ohm shunt at 10A test current.

13-10

Automatic; the ATO will measure the 100

?

-ohm shunt at 100A test current.

Test results will be printed at the end of the
100A test.

13-11

Connect the 500 micro-ohm shunt to the
ATO. Press “ENTER” key to start test.

13-12

Automatic; the ATO will measure the 500

?

-ohm shunt at 10A test current.

13-13

Automatic; the ATO will measure the 500

?

-ohm shunt at 100A test current.

Test results will be printed at the end of
100A test.

13-14

Connect the 1000 micro-ohm shunt to the
ATO. Press “ENTER” key to start test.

13-15

Automatic; the ATO will measure the 1000

?

-ohm shunt at 10A test current.

13-16

Automatic; the ATO will measure the 1000

?

-ohm shunt at 100A test current.

Test results will be printed at the end of the
100A test.

13-17

Connect the 1900 micro-ohm shunt to the
ATO. Press “ENTER” key to start test.

CURRENT: 10.0 AMP
BURN IN: 02 SEC
100

? -OHM TEST

CURRENT: 100.0 AMP
BURN IN: 02 SEC
100

? -OHM TEST

CONNECT 500

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
500

? -OHM TEST

CURRENT: 100.0 AMP
BURN IN: 02 SEC
500

? -OHM TEST

CONNECT 1000

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
1000

? -OHM TEST

CURRENT: 100.0 AMP
BURN IN: 02 SEC
1000

? -OHM TEST

CONNECT 1900

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE