ATO-400P / ATO-600P Operating Procedures

38

7.12 Calibration Test Procedure

This procedure describes the steps to verify the ATO measurement accuracy and generates a
calibration report. Five different calibration shunts are required for this test: 50micro-ohm, 100
micro-ohm, 500 micro-ohm, 1000 micro-ohm, 1900 micro-ohm. These test shunts shall be 0.1%
accuracy or better.

Table 13.0 Calibration Test Procedure

STEP

ACTION

ATO DISPLAY

13-1

Press key # 2 on the START MENU to go
to the SETUP MENU shown at right.

13-2

On the SETUP MENU, press key #4
(NEXT PAGE).

13-3

Press key #2 to select “CALIBRATE
UNIT”.

13-4

Press the “ENTER” key to continue.

13-5

Connect the 50 micro-ohm shunt to the
ATO. Press the “ENTER” key to start test.

13-6

Automatic; the ATO will measure the 50

?

-

ohm shunt at 10A test current.

13-7

Automatic; the ATO will measure the 50

?

-

ohm shunt at 100A test current.
Test results will be printed at the end of the
100A test.

13-8

Connect the 100 micro-ohm shunt to ATO.
Press “ENTER” key to start test.

1. ENTER RECORD ID
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE

1. SET TIME
2. CALIBRATE UNIT

CALIBRATION-SPECIFIC
SHUNTS ARE REQUIRED!
“ENTER” TO CONTINUE

CONNECT 50

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
50

? -OHM TEST

CURRENT: 100.0 AMP
BURN IN: 02 SEC
50

? -OHM TEST

CONNECT 100

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
100

? -OHM TEST