Local (front leveling panel) operation operations – Anritsu 681XXC User Manual
Page 105

Erasing the Power-Offset Tables from Memory
The power-offset tables are stored in non-volatile
memory. A master reset is required to erase the con-
tents of the tables and reprogram them with default
data.
To perform a master reset, proceed as follows:
Step 1
With the 681XXC in standby, press and
hold the
RF OUTPUT ON/OFF
key.
Step 2
Press the
LINE OPERATE/STANDBY
key
to turn the instrument on.
Step 3
When the first menu is displayed (after
the start-up display), release the
RF OUT-
PUT ON/OFF
key.
The contents of non-volatile memory have now been
erased and reprogrammed with default data.
3-72
681XXC OM
LOCAL (FRONT
LEVELING
PANEL) OPERATION
OPERATIONS
NOTE
The master reset function over-
writes all information stored in the
non-volatile memory with default
values. This includes the nine
stored front panel setups and the
table of 2000 frequency/power level
sets used for list sweep mode.