Local (front leveling panel) operation operations – Anritsu 681XXC User Manual
Page 103
Place the signal generator in a fixed power level
mode by pressing the main menu key
LEVEL/ALC
SELECT
At the resulting menu display, press
Level
. The
681XXC is now in fixed (non-swept) power level
mode.
At the Level Menu, press
User Cal
. The User Level
Cal Menu (below) is displayed.
This menu lets you perform the following:
q
Create a power-offset table.
q
Select a measurement frequency range (edit
the start and stop frequency parameters).
q
Select the number of points at which correction
information is to be taken.
q
Apply a power-offset table to the test setup.
First, press the menu soft-key to select the power-
offset table (
User 1
,
User 2
,
User 3
,
User 4
, or
User 5
)
that you wish to create.
Next, set the measurement frequency range by
pressing
Edit Start
or
Edit Stop
to open the start
(
Flo
) or stop (
Fhi
) frequency parameter for editing.
Edit the current frequency using the cursor control
key or rotary data knob or enter a new value using
the keypad and appropriate termination key. When
you have finished setting the open parameter, close
it by pressing its menu edit soft-key or by making
another menu selection.
Then, select the number of frequency points at
which correction information is to be taken by press-
ing
Edit Points
to open the number-of-points
parameter for editing. Edit the current number-of
-points using the cursor control key or rotary data
knob or enter a new value using the keypad and the
STEPS
termination key. (The number of points
3-70
681XXC OM
LOCAL (FRONT
LEVELING
PANEL) OPERATION
OPERATIONS