Performance verification, Test equipment required – Teledyne LeCroy AP034 User Manual
Page 31

Operator’s Manual
922261-00 Rev A
25
Performance Verification
This procedure can be used to verify the warranted characteristics of the AP034 Active Differential 
Probe. You can do the performance verification without removing the instrument covers and 
exposing yourself to hazardous voltages. 
Adjustment should only be attempted if a parameter measured in the Performance Verification 
Procedure is outside of the specified limits. Adjustment should only be performed by qualified 
personnel. 
The recommended calibration interval for the model AP034 Active Differential Probe is one year. 
The complete performance verification procedure should be performed as the first step of annual 
calibration. You can record test results on a photocopy of the Test Record at the end of this 
section. 
Test Equipment Required
Table 4 lists the test equipment and accessories (or their equivalents) that are required for 
performance verification of the AP034 Active Differential Probe. 
This procedure has been developed to minimize the number of calibrated test instruments 
required. Only the parameters listed in boldface in the “Minimum Requirements” column must be 
calibrated to the accuracy indicated. 
Because the input and output connector types may vary on different brands and models of test 
instruments, additional adapters or cables may be required. 