beautypg.com

Teledyne LeCroy HDO 4000 Operators Manual User Manual

Page 114

background image

HDO4000 High Definition Oscilloscope

Run the Recovery Wizard

1. Apply power to the oscilloscope.

2. During the startup process, as soon as you see the message "Starting Acronis Loader... Press F11 for

Acronis Startup Recovery Manager," press the F11 key until the Acronis logo appears momentarily.
The Acronis True Image Home window is displayed.

NOTE: Do not press F11 before you see the "Starting Acronis Loader..." message or you will enter the
boot device selection menu. If a bootmenu dialog box appears, press Cancel or Esc.

3. Select Acronis True Image Home (Full Version).

4. On the Acronis True Image Home page, under options for Recover, select My Disks. The Recovery Wiz-

ard opens.

5. On the Recovery Wizard, under Archive Selection, select the disk archive that has a create date, then

click Next.

6. Under Recovery Method, select Recover whole disks and partitions, then click Next.

7. Under What to Recover, select NTFS (SYSTEM) (C:), then click Next.

8. Under Settings of Partition C, in the top section, Partition location (required), select New Location.The

Partition Destination window opens.

9. Under New Partition Location, select NTFS (SYSTEM) (D:), then click Accept. This returns you to the

Settings of Partition C step. Click Next.

10. A summary window is displayed indicating that Acronis True Image is ready to proceed with Recov-

ering partition C -> D. Click Proceed to start the recovery process.

NOTE: Recovery takes approximately 4 to 15 minutes depending on the version of operating system
that is being restored. The progress is displayed on the screen.

11. When recovery is complete, you will see the message "Recover operation succeeded." Click OK.

12. Click the close button to exit the Acronis window. The oscilloscope will restart and begin installing the

required software.

NOTE: If prompted to restart Windows, select Restart Later.

Reinstall the Oscilloscope Application

1. When the X-Stream DSO Setup Wizard appears, click Next.

2. On the License Agreement page, select I Agree.

3. On the Choose Components page, select the default (installation) and click Install.

NOTE: You may see a Windows Security message indicating that Windows can't verify the publisher of
the driver software. Choose "Install this driver software anyway" and click Install.

4. When the X-Stream installation is complete, choose to Reboot now and click Finish.

Reactivate the F11 Startup Utility

In order for the system recovery wizard to be accessed again from the boot menu, you must reactivate the
F11 startup utility. It is critical to complete these steps after your oscilloscope has restarted.

1. From the Windows Start Menu choose All Programs > Acronis > Acronis True Image Home.

2. On the Acronis True Image Home window, towards the top right, click Tools & Utilities.

106