Solutions – Fairbanks 1129 SERIES Dual Platform Counting Scale User Manual
Page 41
Section 5: Service and Maintenance
08/12
41
51300 Rev. 1
5.2.2. Solutions
For all errors not listed or for help, please call your local Fairbanks Service Representative.
SYMPTOM
CAUSE
SOLUTION
“UL1”
or
“UL2”
• Below
ZERO
• Bind
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the lever’s end at
the bottom of the scale.
Check for proper platform assembly.
Possible damaged load cell.
Call for service.
“OL1”
or
“OL2”
• Over Capacity
• Bind
• Possible damaged
load cell.
• Possible A/D failure.
Check platform to be empty, then press
ZERO
.
Check for platform bind.
Call for service.
Does not return to
“0.0”
• Bind
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the lever’s end at
the bottom of the scale.
Check for proper platform assembly.
“Error3”
• Bind
• Possible damaged
load cell.
• Possible A/D failure.
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the lever’s end at
the bottom of the scale.
Check for proper platform assembly.
Call for service.
Blank display
• Battery pack
discharged.
• No power at outlet
• Batteries dead
• Faulty instrument
Replace or Recharge the battery pack.
Check AC outlet; check circuit box.
Call for service.
“LoBAtt”
• Batteries low
Recharge the battery pack.