Ndir gas analyzer appendix – Teledyne IR-7000 - Non-dispersive infrared series User Manual

Page 85

background image

NDIR

Gas

Analyzer

Appendix


Teledyne Analytical Instruments., - Rev. 3

77

A-3.2 Codes for "s" Type Messages

The codes for "s" type messages are:

C

ODE

R

ETURN

M

ESSAGE

1C 1

st

(ir) Cal gas value

AP

Auto cal Period in hours

1R

chart Range 1 value

2R

chart Range 2 value

3R chart

Range 3 value

4R chart

Range 4 value

1H

Setpoint 2 trigger level

1L

Setpoint 1 trigger level

CD Cal Delay

1X

1

st

oXygen span cal setting

A-3.3 Codes for "c" Type Messages

The codes for "c" type messages are:

C

ODE

M

ODE

N

AME

V

ALUE

PL Panel Lock

1 = lock, 0 = unlock

DB

Display Brightness

0 thru 3 (3 is brightest)

RS

chart Range Selected

0 = Autorange, 1 thru 4

FS

Filter Selected

0 = filter 1

1 = filter 10

2 = filter 100 3 = filter 1000

CT Cal Track/hold

1 = track

0 = hold

TC Timed Cal type

0 = full

1 = Zero only

AE audible

Alarm Enable

0 = disabled 1 = enabled

LM Linear Mode

0 = linear 1 = nonlinear 2 = counts

CS Cal Select

0 = Zero 1 = Span 2 = both

AK

Alarm acKnowledge

1 = acknowledge alarm