LINEARIZER ERROR

The linearizer software
has failed in the self-
diagnostic test while in
use.

Press either UP or DOWN key to
acknowledge the alarm and continue
normal operation.

If the message occurs again, turn the
instrument off for 10 seconds then
power back up.

LINEARIZER FAILURE

The linearizer software
has failed during the
power –up self-diagnostic
test.

To clear the message, turn the
instrument off for 10 seconds, then
power back up.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

CALVALUE CHANGED

The stored calibration
readings have failed in
the continuous self-test.

Press either UP or DOWN key to
acknowledge the alarm and continue
normal operation.

If the message occurs again, turn the
instrument off for 10 seconds then
power back up.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

NO EPROM PRESENT

Some portion of the
permanent memory is not
available to the
microprocessor.

Turn the instrument off for 10 seconds,
then power back up.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

EXTERNAL RAM FAILURE

Some portion of the
addressable memory is
failing initialization self-
test.

Turn the instrument off for 10 seconds,
then power back up.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

NO DATA READY FAILURE

A hardware fault has been
detected in the main
board.

To clear the message, turn the
instrument off for 10 seconds, then
power back up.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

O2 SENSOR FAIL

The digitized oxygen
sensor is significantly out
of range.

Replace the O

2

sensor.

If the message re-occurs, discontinue
use of the instrument and contact LSC.

O2 ZERO FAIL

[CAUSE]

OXYCALS

RESTORED

NEED GOOD CALS

An O

2

zero calibration

was attempted and did not
succeed.

Insufficient cell purge. Increase CAL
DELAY period or purge time.

Check for proper zero gas.

Check sample system for leaks.