beautypg.com

Teledyne 6020 - Photometric analyzer User Manual

Page 90

background image

A-2

Appendix

Models 6020

Teledyne Analytical Instruments

Typical Analytical Performance Specifications:

(

will vary per application

)

Accuracy: ±1% of full scale possible

Noise: Less than ±1%

Drift: Less than 1% per day (source/detector depen-

dent)

Diurnal: Less than 1% per 20

o

F (source/detector

dependent)

Sample Cell: Stainless steel (PVC for Oil in Water) with

Quartz window standard. Other materials
available.

Cell Length: .01 to 40 inches

Flow Rate: 50 to 1500 cc/min

Light Source: Tungsten Lamp, Mercury, Deuterium Arc

Sensitivity: .02 to 3 absorbance units.

Reproducibility: +/-2% of scale or better

Filter Wavelength: 210 to 1000 millimicrons, application depen-

dant

Sample Pressure: Quartz window: 30 psi maximum

Optional: Sapphire 1000psi maximum