28 part i – Teledyne 6020 - Photometric analyzer User Manual
Page 52
3 Operation
Model 6020
3-28 Part I
Teledyne Analytical Instruments
: 4
: 4
: 4
: 4
: 4
Use the UP/DOWN switch again to move the blinking to
APPLICATION
and
Enter
.
:
:
:
:
:
Ñ> 1 2 3
<С
С> 1 2 3
<С
С> 1 2 3
<С
С> 1 2 3
<С
С> 1 2 3
<Ñ
Use the UP/DOWN switch to increment/decrement the range number to
01, 02, 03, or CAL, and
Enter
.
**********
**********
**********
**********
**********
: :1 %
: :1 %
: :1 %
: :1 %
: :1 %
Use the UP/DOWN switch to increment the respective parameters as
desired, and
Enter
to move to the next. On the last field %/ppm Enter to accept
the values and return to range selection menu. (See note below.) Repeat for each
range you want to set.
Note: The ranges must be increasing from low to high, for example,
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then
Range 3 cannot be set to 0–50 % since that makes Range 3
lower than Range 2.
Ranges, alarms, and spans are always set in either percent or ppm units, as
selected by the operator, even though all concentration-data outputs change
from ppm to percent when the concentration is above 9999 ppm.
Note: When performing analysis on a fixed range, if the concentra-
tion rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.
To end the session, send:
st
st
to the analyzer from the computer.
3.8.2 The Curve Algorithm Screen
The Curve Algorithm is a linearization method. It provides from 1 to 9
intermediate points between the ZERO and SPAN values, which can be normal-
ized during calibration, to ensure a straight-line input/output transfer function
through the analyzer.