Photometric Analyzer

Operation 3

Part I 3-5

Teledyne Analytical Instruments

APPLICATION

: Used to define the analysis ranges and application

(gas used).

MODEL

: Displays model number and software version.

OUTPUT_CAL

: 4-20 MA: Adjust 4 and 20 mA output.

The hierarchy of the system menu is shown in figure 3-1.

3.3.1 Setting up an AUTO-CAL

When proper automatic valving is connected, the Analyzer can cycle itself

through a sequence of steps that automatically zero and span the instrument.

Note: Before setting up an AUTO-CAL, be sure you understand the

Zero and Span functions as described in section 3.4, and
follow the precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external

AUTO-CAL control where possible. The internal clock in the
Model 6020 is accurate to 2-3 %. Accordingly, internally sched-
uled calibrations can vary 2-3 % per day.

To setup an

Auto–Cal

cycle:

Choose

System

from the main manu. TheVFD will display five subfunc-

tions.

_

С

_

С

_

С

_

С

_

С

Select

and Enter

-

-

-

-

-

-

-

-

-

-

Use UP/DOWN to blink

Ñ

, and

Enter

. A new screen for

ZERO/SPAN

set appears.

Use UP/DOWN to blink

(or

), then

Enter

again. (You

won’t be able to set

OFF

to

ON

if a zero interval is entered.) A

Span

Every

... (or

Zero Every

...) screen appears.

:

:

:

:

:

:

:

:

:

:

:

:

:

:

: