3 Operation

Model 6020

3-10 Part I

Teledyne Analytical Instruments

3.3.5 The Model Screen

Enter the System menu, select more and Enter. The second screen ap-

pears. Select more again and Enter. In the third screen select MODEL. With

MODEL

blinking,

Enter

. The screen displays the manufacturer, model, and

software version information. Escape to return to the System menu.

3.3.6 Checking Linearity with ALGORITHM

From the

System

Function screen, select

ALGORITHM

, and

Enter

.

:

:

:

:

:

Ñ> 1 2 3 <С

С> 1 2 3 <С

С> 1 2 3 <С

С> 1 2 3 <С

С> 1 2 3 <Ñ

Use the UP/DOWN switch to select the range: 01, 02, or 03. Then

Enter

. (Some ranges may not be available, depending on your application, but

at least range 01 should be).

: 2

: 2

: 2

: 2

: 2

::::: С 10%

С 10%

С 10%

С 10%

С 10%

Enter again.

:

:

:

:

:

Select and

Enter

VERIFY

to check whether the linearization has been

accomplished satisfactorily.

. .

. .

. .

. .

. .

The leftmost digit (under

Dpt

) is the number of the data point being moni-

tored. Use the UP/DOWN switch to select the successive points.

The

INPUT

value is the input to the linearizer. It is the simulated output of

the analyzer. You do not need to actually flow gas.

The

OUTPUT

value is the output of the linearizer. It should be the ACTUAL

concentration of the span gas being simulated.

If the

OUTPUT

value shown is not correct, the linearization must be correct-

ed. Press

ESCAPE

to return to the previous screen. Select and Enter

SET UP

to Calibration Mode screen. (set-up will not work without a PC being connected
to the analyzer)

: