3.3.5 The Model Screen
Enter the System menu, select more and Enter. The second screen ap-
pears. Select more again and Enter. In the third screen select MODEL. With
MODEL
blinking,
Enter
. The screen displays the manufacturer, model, and
software version information. Escape to return to the System menu.
3.3.6 Checking Linearity with ALGORITHM
From the
System
Function screen, select
ALGORITHM
, and
Enter
.
:
:
:
:
:
Ñ> 1 2 3 <С
С> 1 2 3 <С
С> 1 2 3 <С
С> 1 2 3 <С
С> 1 2 3 <Ñ
Use the UP/DOWN switch to select the range: 01, 02, or 03. Then
Enter
. (Some ranges may not be available, depending on your application, but
at least range 01 should be).
: 2
: 2
: 2
: 2
: 2
::::: С 10%
С 10%
С 10%
С 10%
С 10%
Enter again.
:
:
:
:
:
Select and
Enter
VERIFY
to check whether the linearization has been
accomplished satisfactorily.
. .
. .
. .
. .
. .
The leftmost digit (under
Dpt
) is the number of the data point being moni-
tored. Use the UP/DOWN switch to select the successive points.
The
INPUT
value is the input to the linearizer. It is the simulated output of
the analyzer. You do not need to actually flow gas.
The
OUTPUT
value is the output of the linearizer. It should be the ACTUAL
concentration of the span gas being simulated.
If the
OUTPUT
value shown is not correct, the linearization must be correct-
ed. Press
ESCAPE
to return to the previous screen. Select and Enter
SET UP
to Calibration Mode screen. (set-up will not work without a PC being connected
to the analyzer)
: