beautypg.com

Teledyne 6020 - Photometric analyzer User Manual

Page 38

background image

3 Operation

Model 6020

3-14 Part I

Teledyne Analytical Instruments

1.

Enter the System menu:

_

-

2.

Using theUP/DOWN switch, select MORE and press Enter:

The Second System screen appears:

-

-

-

-

3.

Select with the UP/DOWN switch CAL-HOLD-TIMER,

and press the Enter key to access this function menu:

:

3

:

1

The calibration hold time is set on the first row, while the sample hold

time is set on the second row. To select one or the other, use the Right or
Left keys. To modify the time of either timer, use the Up or Down keys.
The time is in the minutes.

3.3.11 Analog 4-20mA Output Calibration

This function will let the operator calibrate the 4 to 20 mA analog output to

match the display reading. A DMM configure as a DC ammeter is needed. The
DMM should be connected across the output terminals of the 4 to 20 mA output
to monitor the output current. To enter the 4 to 20 mA output adjust function,
you must:

1.

Enter the System menu:

_

-

2.

Using the Right or Left arrow keys, select MORE and press

Enter. The second System screen appears:

-

-

-

-