beautypg.com

18 part i – Teledyne 6020 - Photometric analyzer User Manual

Page 42

background image

3 Operation

Model 6020

3-18 Part I

Teledyne Analytical Instruments

The zeroing process will automatically conclude when the output is within

the acceptable range for a good zero. Then the analyzer automatically returns to
the

Analyze

mode.

3.4.1.2

Manual Mode Zeroing

Enter the

Zero

function. The screen that appears allows you to select

between automatic or manual zero calibration. Use the UP/DOWN switch
between

AUTO

and

MAN

zero settling. Stop when

MANUAL

appears, blinking,

on the display.

:

:

:

:

:

Enter

to begin the zero calibration. After a few seconds the first of three

zeroing screens appears. The number in the upper left hand corner is the first-
stage zero offset. The microprocessor samples the output at a predetermined
rate.

####.##

####.##

####.##

####.##

####.##

:2048 С

:2048 С

:2048 С

:2048 С

:2048 С

The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero

and F–Zero, use the UP/DOWN

SWITCH

to adjust displayed

Zero adj:

value

as close as possible to zero. Then,

Enter

.

S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto

Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as

Slope=

a value in parts per million per

second (ppm/s).

####.##

####.##

####.##

####.##

####.##

=#.### С

=#.### С

=#.### С

=#.### С

=#.### С

Once zero settling completes, the information is stored in the analyzer’s

memory, and the instrument automatically returns to the

Analyze

mode.

3.4.1.3 Detector Failure

Detector failure in the 6020 is usually associated with inability to zero the

instrument with a reasonable voltage differential between the reference and
measure voltages. If this should ever happen, the 6020 system alarm trips, and
the LCD displays a failure message.