Photometric Analyzer

Operation 3

Part I 3-23

Teledyne Analytical Instruments

3.6

The Range Select Function

The

Range

function allows you to manually select the concentration range

of analysis (

MANUAL

), or to select automatic range switching (

AUTO

).

In the

MANUAL

screen, you are further allowed to define the high and low

(concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must

be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.

3.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Enter the

Range

function to start the Range function.

:

:

:

:

:

If above screen displays, use the UP/DOWN switch to Select

MANUAL

,

and

Enter.

Ñ> 1 2 3 04<С

С> 1 2 3 04<С

С> 1 2 3 04<С

С> 1 2 3 04<С

С> 1 2 3 04<Ñ

NOTE: Oil in Water applications require single range (01) measurement.

Use the UP/DOWN switch to select the range: 01, 02, 03, or 04. Then

Enter

.

:

:

:

:

:

: С 5%

: С 5%

: С 5%

: С 5%

: С 5%

The high-end of the range field should blink first. Use UP/DOWN switch

to change the value of the field, Enter to move to the low-end of the range field.

Escape

to return to the previous screen to select or define another range.

Enter

to return the to the

Analyze

function.