Teledyne 6020 - Photometric analyzer User Manual
Page 47
Photometric Analyzer
Operation 3
Part I 3-23
Teledyne Analytical Instruments
3.6
The Range Select Function
The
Range
function allows you to manually select the concentration range
of analysis (
MANUAL
), or to select automatic range switching (
AUTO
).
In the
MANUAL
screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
3.6.1 Manual (Select/Define Range) Screen
The Manual range-switching mode allows you to select a single, fixed
analysis range. It then allows you to redefine the upper and lower limits, for the
range.
Enter the
Range
function to start the Range function.
:
:
:
:
:
If above screen displays, use the UP/DOWN switch to Select
MANUAL
,
and
Enter.
Ñ> 1 2 3 04<С
С> 1 2 3 04<С
С> 1 2 3 04<С
С> 1 2 3 04<С
С> 1 2 3 04<Ñ
NOTE: Oil in Water applications require single range (01) measurement.
Use the UP/DOWN switch to select the range: 01, 02, 03, or 04. Then
Enter
.
:
:
:
:
:
: С 5%
: С 5%
: С 5%
: С 5%
: С 5%
The high-end of the range field should blink first. Use UP/DOWN switch
to change the value of the field, Enter to move to the low-end of the range field.
Escape
to return to the previous screen to select or define another range.
Enter
to return the to the
Analyze
function.