Teledyne 6020 - Photometric analyzer User Manual

Page 28

background image

3 Operation

Model 6020

3-4 Part I

Teledyne Analytical Instruments

Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)

Zero

. Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine whether

each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.

Range.

Used to set up four analysis ranges that can be switched

automatically with autoranging or used as individual fixed
ranges.

Any function can be selected at any time in the analyze mode (unless

password restrictions apply). The order as presented in this manual is
appropriate for an initial setup.

Each of these functions is described in greater detail in the following proce-

dures. The VFD screen text that accompanies each operation is reproduced, at
the appropriate point in the procedure, in a

type style.

3.3

The System Function

The subfunctions of the

System

function are described below. Specific

procedures for their use follow the descriptions:

Dig_Filt

: Adjust how much digital filtering should be on the

signal

SELF-TEST

: Performs a self-diagnostic test to check the integrity

of the power supplies, outputs, detector signal and preamplifier.

PWD

: Login security system for accessing to the setup functions.

LOGOUT

: Prevents an unauthorized tampering with analyzer

settings.

AUTOCAL

: Set the automatic calibrated timer schedule for Zero

and Span cycling.

FILSOL

: Select Span/Zero flag (filter) or Span/Zero solenoid

valve for calibration method.

TRACK

: Set the system reading to be held or followed by the

concentration “gas or filter” during calibration.

CAL-HOLD-TIMER

: Set the timing for calibration holding and

timing for the sample reading after return to analyze mode.

ALGORITHM

: Linearize the output for nonlinear characteristic.