Teledyne 6020 - Photometric analyzer User Manual
Page 41
Photometric Analyzer
Operation 3
Part I 3-17
Teledyne Analytical Instruments
3.4.1 Zero Cal
The
Zero
function on the main menu is used to enter the zero calibration
function. Zero calibration can be performed in either the automatic or manual
mode.
Make sure the zero fluid is flowing to the instrument. If you get a
CELL
CANNOT BE BALANCED
message while zeroing skip to section 3.4.1.3.
3.4.1.1
Auto Mode Zeroing
Observe the precautions in sections 3.4 and 3.4.1, above. Enter the zero
function mode. The screen allows you to select whether the zero calibration is to
be performed automatically or manually. Use the UP/DOWN switch to toggle
between
AUTO
and
MAN
zero settling. Stop when
AUTO
appears, blinking, on
the display.
:
:
:
:
:
Enter
to begin zeroing.
####.##
####.##
####.##
####.##
####.##
=#.### С
=#.### С
=#.### С
=#.### С
=#.### С
The beginning zero level is shown in the upper left corner of the display. As
the zero reading settles, the screen displays and updates information on
Slope=
in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a coarse zero, shown in
the lower right corner of the screen as
C—Zero
, for 3 min, and then does a fine
zero, and displays
F—Zero
, for 3 min.
Then, and whenever Slope is less than 0.01 for at least 12 sec, instead of
Slope you will see a countdown:
9 Left
,
8 Left
, and so fourth. These are
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to
Analyze
. Software zero is indicated by
S–Zero
in the lower right corner.
NOTE:
In a Oil/Water sampling system, when performing a sched-
uled zero, instrument will go to span mode automatically (when span flag
option has been purchased).
####.##
####.##
####.##
####.##
####.##