3 Operation
Model 6020
3-26 Part I
Teledyne Analytical Instruments
3.7
The Analyze Function
Normally, all of the functions automatically switch back to the
Analyze
function when they have completed their assigned operations. The
Escape
key
in many cases also switches the analyzer back to the
Analyze
function.
The
Analyze
function screen shows the impurity concentration in the first
line, and the range in the second line. In the lower right corner, the abbreviation
Anlz
indicates that the analyzer is in the
Analyze
mode. If there is an * before
the
Anlz
, it indicates that the range is linearized.
1.95 %
1.95 %
1.95 %
1.95 %
1.95 %
1: С10 *
1: С10 *
1: С10 *
1: С10 *
1: С10 *
If the concentration detected is overrange, the first line of the display blinks
continuously.
3.8
Programming
CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the facto-
ry to the users application specification. These func-
tions should only be reprogrammed by trained,
qualified personnel.
To program, you must:
1. Enter the password, if you are using the analyzer’s password
protection capability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 2
Installation for details). Send the rp command to the analyzer.
3. Enter the
System
menu.
_
-
_
-
_
-
_
-
_
-
Use the UP/DOWN switch to blink
MORE
, then
Enter
.
-
-
-
-
-
-
-
-
-
-
Select MORE and ENTER one more time
: 4