Teledyne 6020 - Photometric analyzer User Manual

Page 57

background image

Part I: Control Unit

Maintenance 4

Part I: 4-3

Teledyne Analytical Instruments

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed

>3

Call factory for information

Detector

0

OK

1

Failed (open filament, short to ground, no
power.)

2

Unbalance (deterioration of filaments, blocked
tube)

4.3

Major Internal Components

All internal components are accessed by unbolting and swinging open

the front cover, as described earlier. The major internal component locations
are shown in Figure 4-2, the cell block is illustrated in Figure 3-2, and the
fuse receptacle is shown in Figure 3-3

The 6020 contains the following major internal components:

Customer Interface PCB (Power Supply on bottom surface)

Preamp PCB (Contains Microprocessor)

Front Panel PCB (Contains Displays)

5 digit LED meter
2 line, 20 character, alphanumeric, VFD display

See the drawings in the Drawings section in back of this manual

for details.

For Optical/Detector Alignments, refer to parts II or III of this manual