beautypg.com

Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 68

background image

A-2

Appendix

Model Ultra Trace 3000

Teledyne Analytical Instruments

Power: Universal power supply 85-250 V ac, at

47-63 Hz.

Operating Temperature: 5-40 °C

Accuracy: ±2% of full scale for all ranges except 0-250

ppb range, at constant temperature.

For 0-250 ppb range accuracy is as follows:

±5 ppb at constant temperature.

±10 ppb over temperature variation of 20° -
30°C.

±15 ppb over temperature variation of 15° -
35°C.

All accuracy specifications are contingent
upon the completion of zero ans span calibra-
tion.

All accuracy is established at constant pres-
sure and equilibrium has been established.

Analog outputs: 0-1 V dc percent-of-range,

0-1 V dc range ID.
4-20 mA dc (isolated) percent-of-range,
4-20 mA dc (isolated) range ID.

Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep (6.96

in high, 8.7 in wide, 12.2 in deep).