4 Operation

Model Ultra Trace 3000

4-10

Teledyne Analytical Instruments

status of each function is displayed on the screen, either as

OK or as a num-

ber between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)

The self diagnostics are run automatically by the analyzer whenever the

instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation:

Press the

System

button to start the

System

function.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the < > arrow keys to blink

More, then press

Enter

.

Version Self—Test

Neg-N Filter-5

Use the < > arrow keys again to move the blinking to the

Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC

Testing Preamp — 83

During preamp testing there is a countdown in the lower right corner of

the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK

Preamp: 3

The module is functioning properly if it is followed by

OK. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
for number-code information. The results
screen alternates for a time with:

Press Any Key

To Continue...

Then the analyzer returns to the initial System screen.