3 Installation

Model Ultra Trace 3000

3-2

Teledyne Analytical Instruments

Figure 3-1: Front Panel of the Model Ultra Trace 3000

All operator controls are mounted on the control panel, which is hinged

on the left edge and doubles as the door that provides access to the sensor
and cell block inside the instrument. The door is spring loaded and will
swing open when the button in the center of the latch (upper right corner) is
pressed all the way in with a narrow gauge tool (less than 0.18 inch wide),
such as a small hex wrench or screwdriver Allow clearance for the door to
open in a 90-degree arc of radius 7.125 inches. See Figure 3-2.

Provide 1 inch of air clearance at the bottom and sides of the instrument

for proper air flow for TEC fun.

Failure to provide such a clearance will cause damage to the TEC

system. Do not restrict this openings.

Figure 3-2: Required Front Door Clearance

2.0

1.0

Hinge

Latch

Analyze

System

Span

Zero

Alarms

Range

Standby

Enter

Escape

Set Flow

ULTRA TRACE 3000

ULTRA TRACE ANALYZER

6.7”

10”